The HI6000-2 module measures pH, ORP, specific ions and temperature.
Measurement
- Choice of Measurement Unit
- pH (HI6000-1, HI6000-2) – pH, mV
- ORP (HI6000-1, HI6000-2) – mV, Rel.mV
- ISE (HI6000-2 only) – ppt, ppm, ppb, g/L, mg/L, μg/L, mg/mL, μg/mL, M, mol/L, mmol/L, %w/v, user defined
- Reading modes:
- Direct and direct/autohold
- Known Addition, Known Subtraction, Analyte Addiiton,
- Analyte Subtration (HI6000-2)
Calibration
- pH calibration using
- Up to five Hanna Instruments pH buffers (pH 1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01 and 12.45)
- Up to five custom buffers
- mV calibration using a single point to calibrate offset.
- ISE calibration using up to five nominal standard values and/or up to five custom solutions (user supplied)
Recommended probes
For pH measurements, Hanna recommends the HI1131B double junction combination pH electrode, together with HI7662-TW temperature probe for use with these modules.
HI1131B is a glass body, double junction, refillable pH electrode with an indicating sensor made of high temperature (HT) glass. The double junction reference and HT glass design allow the electrode to be used in a
wide variety of applications.
Probe connection to the unit is secured through a galvanically isolated BNC connection.
HI7662-TW stainless steel temperature probe allows the meter to automatically temperature compensate (ATC) pH measurements.
Specifications
Specification | HI6000-2 |
pH Range | -2.0 to 20.0 pH; -2.00 to 20.00 pH; -2.000 to 20.000 pH |
pH Resolution | 0.1 pH; 0.01 pH; 0.001 pH |
pH Accuracy | ±0.1 pH; ±0.01 pH; ±0.002 pH (±1 last significant digit) |
Temperature Compensation | Automatic or manual |
Calibration Points | Up to 5 |
Calibration Type | Automatic; Semiautomatic; Manual |
Standard Buffers | Hanna and NIST pH 1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45 |
Custom Buffers | Up to 5 |
Custom Group | Up to 5 |
1st Calibration Point | Offset or point (user setting) |
Isopotential Point | -2.000 to 20.000 pH |
mV Range | -2000.0 mV to 2000.0 mV |
mV Resolution | 1 mV; 0.1 mV |
mV Accuracy | ±0.2 mV ±1 last significant digit |
mV Calibration | Single point offset, ±2000.0 mV |
ISE Range | 1.0×10⁻⁵ to 300.0 ppt (g/L or mg/mL); 5.0×10⁻³ to 1.0×10⁵ ppm (mg/L or μg/mL); 1.0 to 5.0×10⁷ ppb (μg/L); 1.0×10⁻⁷ to 10.0 M (mol/L); 1.0×10⁻⁴ to 1.0×10⁴ mmol/L; 1.0×10⁻⁶ to 60.0 % w/v; 5.0×10⁻⁷ to 5.0×10⁷ user |
ISE Resolution | 1, 2, 3 significant digits |
ISE Accuracy | ±0.5% (monovalent ions); ±1% (divalent ions) |
ISE Calibration Points | Up to 5 |
ISE Calibration Type | All standards Standard group |
Standards | 7 standard solutions available for each concentration unit |
Custom Standards | Up to 5 |
Custom Group | Up to 5 |
Temperature Range | -20.0 to 120.0 °C; -4.0 to 248.0 °F; 253.2 to 393.2 K |
Temperature Resolution | 0.1 °C; 0.1 °F ; 0.1 K |
Temperature Accuracy | ±0.2 °C; ±0.4 °F; ±0.2 K |
Temperature Calibration | Single point, adjustable |