Module for pH, ORP, specific ions and temperature – HI6000-2

The HI6000-2 module measures pH, ORP, specific ions and temperature.

SKU: HI6000-2 Categories: , ,

The HI6000-2 module measures pH, ORP, specific ions and temperature.

Measurement

  • Choice of Measurement Unit
    • pH (HI6000-1, HI6000-2) – pH, mV
    • ORP (HI6000-1, HI6000-2) – mV, Rel.mV
    • ISE (HI6000-2 only) – ppt, ppm, ppb, g/L, mg/L, μg/L, mg/mL, μg/mL, M, mol/L, mmol/L, %w/v, user defined
  • Reading modes:
    • Direct and direct/autohold
    • Known Addition, Known Subtraction, Analyte Addiiton,
    • Analyte Subtration (HI6000-2)

Calibration

  • pH calibration using
    • Up to five Hanna Instruments pH buffers (pH 1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01 and 12.45)
    • Up to five custom buffers
  • mV calibration using a single point to calibrate offset.
  • ISE calibration using up to five nominal standard values and/or up to five custom solutions (user supplied)

Recommended probes

For pH measurements, Hanna recommends the HI1131B double junction combination pH electrode, together with HI7662-TW temperature probe for use with these modules.

HI1131B is a glass body, double junction, refillable pH electrode with an indicating sensor made of high temperature (HT) glass. The double junction reference and HT glass design allow the electrode to be used in a
wide variety of applications.

Probe connection to the unit is secured through a galvanically isolated BNC connection.

HI7662-TW stainless steel temperature probe allows the meter to automatically temperature compensate (ATC) pH measurements.

 

Specifications

Specification HI6000-2
pH Range -2.0 to 20.0 pH; -2.00 to 20.00 pH; -2.000 to 20.000 pH
pH Resolution 0.1 pH; 0.01 pH; 0.001 pH
pH Accuracy ±0.1 pH; ±0.01 pH; ±0.002 pH (±1 last significant digit)
Temperature Compensation Automatic or manual
Calibration Points Up to 5
Calibration Type Automatic; Semiautomatic; Manual
Standard Buffers Hanna and NIST pH 1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45
Custom Buffers Up to 5
Custom Group Up to 5
1st Calibration Point Offset or point (user setting)
Isopotential Point -2.000 to 20.000 pH
mV Range -2000.0 mV to 2000.0 mV
mV Resolution 1 mV; 0.1 mV
mV Accuracy ±0.2 mV ±1 last significant digit
mV Calibration Single point offset, ±2000.0 mV
ISE Range 1.0×10⁻⁵ to 300.0 ppt (g/L or mg/mL); 5.0×10⁻³ to 1.0×10⁵ ppm (mg/L or μg/mL); 1.0 to 5.0×10⁷ ppb (μg/L); 1.0×10⁻⁷ to 10.0 M (mol/L); 1.0×10⁻⁴ to 1.0×10⁴ mmol/L; 1.0×10⁻⁶ to 60.0 % w/v; 5.0×10⁻⁷ to 5.0×10⁷ user
ISE Resolution 1, 2, 3 significant digits
ISE Accuracy ±0.5% (monovalent ions); ±1% (divalent ions)
ISE Calibration Points Up to 5
ISE Calibration Type All standards
Standard group
Standards 7 standard solutions available for each concentration unit
Custom Standards Up to 5
Custom Group Up to 5
Temperature Range -20.0 to 120.0 °C; -4.0 to 248.0 °F; 253.2 to 393.2 K
Temperature Resolution 0.1 °C; 0.1 °F ; 0.1 K
Temperature Accuracy ±0.2 °C; ±0.4 °F; ±0.2 K
Temperature Calibration Single point, adjustable