[:lt]Modulinis daugiaparametrinis stalinis matuoklis su pH/ORP/ISE + poliarografiniu DO rinkiniu - HI6542P-02[:en]Modular multiparameter benchtop meter with pH/ORP/ISE + polarographic DO kit - HI6542P-02[:] — Modular multiparameter benchtop meter with pH/ORP/ISE + polarographic DO kit - HI6542P-02 [:lt]pH, ORP, specifinių jonų ir temperatūros modulis - HI6000-2[:en]Module for pH, ORP, specific ions and temperature - HI6000-2[:] Module for pH, ORP, specific ions and temperature - HI6000-2 —

pH, ORP and temperature module – HI6000-1

165.00  + PVM

The HI6000-1 module measures pH, ORP and temperature.

SKU: HI6000-1 Categories: ,

Available on backorder - This item may take 2-3 weeks to deliver.

The HI6000-1 module measures pH, ORP and temperature.

Measurement

  • Choice of Measurement Unit
    • pH (HI6000-1, HI6000-2) – pH, mV
    • ORP (HI6000-1, HI6000-2) – mV, Rel.mV
    • ISE (HI6000-2 only) – ppt, ppm, ppb, g/L, mg/L, μg/L, mg/mL, μg/mL, M, mol/L, mmol/L, %w/v, user defined
  • Reading modes:
    • Direct and direct/autohold
    • Known Addition, Known Subtraction, Analyte Addiiton,
    • Analyte Subtration (HI6000-2)

Calibration

  • pH calibration using
    • Up to five Hanna Instruments pH buffers (pH 1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01 and 12.45)
    • Up to five custom buffers
  • mV calibration using a single point to calibrate offset.
  • ISE calibration using up to five nominal standard values and/or up to five custom solutions (user supplied)

Recommended probes

For pH measurements, Hanna recommends the HI1131B double junction combination pH electrode, together with HI7662-TW temperature probe for use with these modules.

HI1131B is a glass body, double junction, refillable pH electrode with an indicating sensor made of high temperature (HT) glass. The double junction reference and HT glass design allow the electrode to be used in a
wide variety of applications.

Probe connection to the unit is secured through a galvanically isolated BNC connection.

HI7662-TW stainless steel temperature probe allows the meter to automatically temperature compensate (ATC) pH measurements.

 

Specifications

Specification HI6000-1 
pH Range -2.0 to 20.0 pH; -2.00 to 20.00 pH; -2.000 to 20.000 pH
pH Resolution 0.1 pH; 0.01 pH; 0.001 pH
pH Accuracy ±0.1 pH; ±0.01 pH; ±0.002 pH (±1 last significant digit)
Temperature Compensation Automatic or manual
Calibration Points Up to 5
Type Automatic; Semiautomatic; Manual
Standard Buffers Hanna and NIST pH 1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45
Custom Buffers Up to 5
Custom Group Up to 5
1st Calibration Point Offset or point (user setting)
Isopotential Point -2.000 to 20.000 pH
mV Range -2000.0 mV to 2000.0 mV
mV Resolution 1 mV; 0.1 mV
mV Accuracy ±0.2 mV ±1 last significant digit
mV Calibration Single point offset, ±2000.0 mV
Temperature Range -20.0 to 120.0 °C; -4.0 to 248.0 °F; 253.2 to 393.2 K
Temperature Resolution 0.1 °C; 0.1 °F ; 0.1 K
Temperature Accuracy ±0.2 °C; ±0.4 °F; ±0.2 K
Temperature Calibration Single point, adjustable

Related products