Setup screen
Our extensive setup screen features a host of configurable options such as time, date, temperature units and language for help screens and guides.
Calibration
pH calibration features detailed CAL Check™ messages. Users are guided through the calibration procedure with step-by-step on-screen instructions.
GLP
Comprehensive GLP functions are directly accessible by pressing the GLP key. Calibration data, including: date, time, offset, slope and buffers used in the last calibration can be reviewed.
- Ergonomic, rugged, waterproof (IP67) design
- pH CAL Check™
- Alerts user to potential problems during calibration including when to clean electrode and possible buffer contamination.
- Displays overall probe condition based on offset and slope after calibration
- Automatic Temperature Compensation
- Five-point pH calibration with seven standard buffers and five custom buffers
- Log-on-demand
- Store measurement data at the press of a button
- AutoHold – freezes reading on display upon stabilization
- GLP features
- GLP data including date, time, pH calibration buffers, offset and slope
- Calibration timeout to alert the user at a defined interval when the calibration has expired
- Multiple language selection
- 200 hour battery life with battery level displayed on measurement screen
- Menu driven for ease of use with virtual-key extended functionality
- Contextual help at the press of a button
- Backlit, graphic LCD
- Connectivity
- PC connectivity via a sealed opto-isolated micro USB and HI92000 software
Supplied with
- HI-12963 titanium body pH electrode
- HI-7004M
- HI-7007M
- 2 x sachets of electrode cleaning solution
- Rugged carrying case
- 100mL plastic beaker (2)
- HI-92000 PC software
- HI-920015 micro USB cable
- Batteries
- Instruction manual
- Quality certificate
Specifications
pH Range | -2.0 to 20.0 pH; -2.00 to 20.00 pH; -2.000 to 20.000 pH | |
pH Resolution | 0.1 pH; 0.01 pH; 0.001 pH | |
pH Accuracy | ±0.1 pH; ±0.002 pH | |
pH Calibration | up to five-point calibration, seven standard buffers available (1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45) and five custom buffers | |
Temperature Compensation | automatic or manual from -20.0 to 120.0°C | |
mV Range | ±2000 mV | |
mV Resolution | 0.1 mV | |
mV Accuracy | ±0.2 mV | |
Relative mV Offset Range | ±2000 mV | |
Temperature Range | -20.0 to 120.0°C | |
Temperature Resolution | 0.1°C | |
Temperature Accuracy | ±0.4°C | |
Additional Specifications: | ||
pH Probe | HI-12963 titanium body, pH electrode with internal temperature sensor, quick DIN connector and 1m cable | |
Slope Calibration | from 80 to 110% | |
Log-on-demand | 200 samples (100 each pH/mV range) | |
PC Connection | opto-isolated USB with HI-92000 software and micro USB cable | |
Input Impedance | 10¹²Ω | |
Battery Type/Life | 1.5 C AA batteries (4) / approximately 200 hours of continuous use without backlight (50 hours with backlight) | |
Auto-off | user selectable; 5, 10, 30, 60 min, disabled | |
Environment | 0 to 50°C; RH 100 % IP67 | |
Dimensions/Weight | 185 x 93 x 35.2 mm / 400g |
Additional Informations
Industry | Aquaculture and aquariums, Environment, Industrial |
---|---|
Parameter | ORP, pH |