pH Analog Controller with Self-Diagnostic Test – HI8710

708.50 € + PVM

HI8710 is a panel mounted pH controller with self-diagnostic test capabilities. 

SKU: HI8710 Category:

Available on backorder - This item may take 2-3 weeks to deliver.

HI8710 is a panel mounted pH controller with self-diagnostic test capabilities. Users can set: the setpoint for acid or alkaline dosage, the tolerance of the setpoint before an alarm is activated, the dosage mode: automatic, continuous on or OFF and the over dosage control by setting the overtime dosage knob.

  • Hanna’s exclusive Fail Safe Fail Safe Alarm System protects against problems caused by power supply failure or signal interruption, which are typical risks in industrial environments.
  • Temperature compensation ensures fast, accurately compensated readings even during sudden temperature fluctuations.
  • Backlit, LCD display for easy viewing.     

 

Details

HI8710  have  been  designed  for  easy  and  fast  installation,and  are  provided  with  membrane  keypad  on  the  front  panel, large display and auto diagnostic functions. All connections are made through screw terminals on the rear panel.

Each model accepts either a direct input from a pH or ORP electrode or from a transmitter through 4-20mA  input. 

The  instrument  also provides  ±5V  power  output  and  input  terminals  for  amplified electrodes. Moreover,  you  can  choose  the  output  configuration  for  connecting  a recorder or a PLC, between 0-20 or 4-20 mA.

HI8710 FEATURES/BENEFITS:

Fail Safe Alarm System

  • Hanna’s exclusive Fail Safe Alarm System protects against problems caused by power supply failure or signal interruption, which are typical risks in industrial environments. An alarm condition is indicated by a red LED, located directly on the front panel of the controller.

Overtime Dosing Protection

  • The HI8711 incorporates adjustable overtime dosing protection from 5 to 60 minutes. If dosing exceeds selected time, the alarm will be triggered and the dosing contact will deactivate. This feature can be activated or deactivated.

Output and Input Connections

  • Each model accepts either a direct input from a pH or ORP electrode or from a transmitter through 4-20 mA input. The instrument also provides ±5V power output and input terminals for amplified electrodes. Moreover, you can choose the output configuration for connecting a recorder or a PLC, between 0-20 or 4-20 mA

Self-Diagnostic Test

  • Built-in autodiagnostic functions enable the user to check and troubleshoot any malfunctions. The functions are made via front panel keys to isolate the cause of malfunction whether it is due to pH electrode contamination, internal offset circuit or the amplifier circuit.

Matching Pin Connection

  • Supports potential matching pin connection to avoid typical problems caused by grounding loop current, such as progressive damage of the electrode, fluctuating measurements, and poor process regulation.

Robust Design

  • A removable, transparent splash-proof cover protects the front panel.

 

 
pH Range 0.00 to 14.00 pH
pH Resolution 0.01 pH
pH Accuracy ±0.02 pH (0 to 100 °C); ±0.05 pH (-20 to 0 °C); ±0.5% (input from transmitter)
pH Calibration offset: ±2 pH with OFFSET trimmer; slope: 80 to 110% with SLOPE trimmer
Temperature Compensation fixed or automatic with Pt100, from -20 to 100°C (-4 to 212°F)
Alarm Relay Set Point 1, isolated, 2 A, max 240 V, resistive load, 1000000 strokes (not fuse protected)
Alarm Relay Range 0.2 to 3.00 pH
Dosing Control OFF/AUTO/ON with selection switch
Over Dosing Control adjustable, from 5 min to 60 min with knob or disable by wire strap – on rear panel
Consent Relay 1, isolated, 2 A, max 240 V, resistive load, 1000000 strokes (not fuse protected)
Analog Ouput Signal 0-20 mA or 4-20 mA (isolated)
Backlight continuous on
Voltage Output ±5 Vcc; 150 mA max load for amplified electrodes
Enclosure flame retardant ABS body and front panel; transparent splash-proof front cover
Panel Cutout 141 x 69 mm (5.6 x 2.7’’)
Setpoint 1, isolated, 2 A, max 240 V, resistive load, 1000000 strokes (not fuse protected); range: 0.00 to 14.00 pH
Input Impedance high impedance 10¹² Ohm; reference and matching pin inputs are available; 4-20 mA
Power Supply 230 Vac, 60/50 Hz
Environment -10 to 50°C (14 to 122°F); RH max 95% non-condensing
Weight 1 kg (2.2 lb.)
Ordering Information The HI8710 is supplied complete with mounting brackets and instructions.