[:lt]Nešiojamasis pH, šarmingumo, laisvo ir bendro chloro ir cianuro rūgšties fotometras - HI971044[:en]pH, Alkalinity, Free and Total Chlorine and Cyanuric Acid Portable Photometer - HI971044[:] — pH, Alkalinity, Free and Total Chlorine and Cyanuric Acid Portable Photometer - HI971044 [:lt]HI981914 Pool Line Profesionalus vandeniui atsparus nešiojamas pH/ORP matuoklis[:en]HI981914 Pool Line Professional Waterproof Portable pH/ORP meter[:] HI981914 Pool Line Professional Waterproof Portable pH/ORP meter —

HI981954 Pool Line Multiparameter Waterproof Portable pH/ORP/EC/Temperature meter

1,171.50  + PVM

Pool Line Multiparameter Waterproof Portable pH/ORP/EC/Temperature meter

SKU: HI981954 Categories: , Tag:

Available on backorder - This item may take 2-3 weeks to deliver.

The HI981954 is a waterproof portable logging multiparameter meter that monitors up to 9 different water quality parameters. It’s multi-sensor  probe allows for the measurement of key parameters including pH, ORP, conductivity, and temperature. The probe transmits readings digitally to the meter, where data points can be displayed and logged. The complete system is simple to setup and easy to use.

The meter has a built-in pressure transducer for barometric pressure readings.

  • One probe with field replaceable pH/ORP, EC and temperature sensors
  • Auto-ranging for EC and TDS readings
  • Display up to 11 measurements simultaneously

pH Features

  • Calibration- up to a three-point calibration with five standard buffers and one custom buffer available
  • pH in mV option – useful for diagnostics
  • GLP data – offset, slope, date, time and buffers used
  • Automatically temperature compensated readings
  • pH or pH/ORP field replaceable sensors
    • Gel filled and maintenance free
    • Double junction for reduced contamination of reference cell

EC/TDS/Resistivity Features

  • Calibration – single-point calibration from six standards
  • Temperature compensation
    • Automatic Temperature Compensation
    • Configurable temperature coefficient range from 0.00 to 6.00%/°C
    • Choice of reference temperatures at 20 or 25°C
    • Absolute conductivity can be displayed along with the temperature compensated value
  • Auto-ranging
  •  Salinity readings
    • Practical Salinity Scale (PSU) based on conductivity calibration

Backlit Graphic LCD Display

  • The HI981954 features a backlit graphic LCD with on-screen help and the capability to display up to nine parameters simultaneously. The graphic  display allows for the use of virtual keys to provide for an intuitive user interface.

Automatic Temperature Compensation

  • Integrated temperature sensor allows for automatic temperature compensation of pH and conductivity measurements.

Waterproof Protection

  • The meter is enclosed in an IP67 rated waterproof casing and can withstand immersion in water at a depth of 1 m for up to 30 minutes. The probe  features an IP68 rating for continuous immersion in water.

Dedicated Help Key

  • Contextual help is always available through a dedicated “HELP” key. Clear tutorial messages and directions are available on-screen to quickly and  easily guide users through setup and calibration. The help information displayed is relative to the setting/option being viewed.

GLP Data

  • HI981954 includes a GLP feature that allows users to view calibration data and calibration expiration information at the touch of a key. Calibration  data includes date, time, buffers/standards used for calibration, and slope characteristics.

PC Connectivity

  • Logged data can be transferred to a Window’s compatible PC with the included HI920015 micro USB cable and HI9298194 software.

Quick Connect Digital Probe

  • The HI7698195 probe features a Quick Connect DIN connector that makes a waterproof connection with the meter.

Long Battery Life

  • The display of the meter has a battery icon indicator to show the remaining power. The meter uses four 1.5V AA batteries that provide up to 360  hours of battery life.

Color Coded, Field Replaceable Sensors

  • Sensor replacement is quick and easy with field replaceable, screw type connectors that are color coded for easy sensor port identification

Data Logging

  • The HI981954 allows users to store up to 45,000 continuous or log-on-demand samples with logging intervals from one second to three hours.

Standard or Quick Calibration

  • Quick Calibration provides a speedy, single point calibration for pH and conductivity. Standard pH calibration options are available for calibrating up  to three points from a selection of five standard buffers and one custom buffer. Conductivity calibration is a single point from six standard  selections or one custom standard.

Intuitive Keypad

  • The fitted rubber keypad has dedicated keys for power, backlight, up/down arrows, help and alphanumeric characters. The meter also features two  virtual soft keys that navigate the user through the configuration of each parameter, meter setup, and logging of data. The interface is intuitive for  any user’s level of experience.

Auto-sensor Recognition

  • The probe and meter automatically recognize the sensors that are connected. Any ports not used on the probe will not have the parameter  displayed or be configurable.

Rugged Custom Carrying Case

  • The HI981954 meter, probe, and all accessories are supplied in a rugged carrying case designed to provide years of use. The inside compartment of the carrying case is thermoformed to securely hold and protect all of the components.

Specifications

pH / mV Range
Resolution
Accuracy
Calibration
0.00 to 14.00 pH / ±600.0 mV
0.01 pH / 0.1 mV
±0.02 pH / ±0.5 mV
automatic one, two, or three points with automatic recognition of five standard buffers
(pH 4.01, 6.86, 7.01, 9.18, 10.01) or one custom buffer
ORP Range
Resolution
Accuracy
Calibration
±2000.0 mV
0.1 mV
±1.0 mV
automatic at one custom point (relative mV)
EC Range
Resolution   

Accuracy

Calibration

0 to 200 mS/cm (absolute EC up to 400 mS/cm)
manual: 1 µS/cm; 0.001 mS/cm; 0.01 mS/cm; 0.1 mS/cm; 1 mS/cm; automatic: 1 µS/cm from 0 to 9999 µS/cm; 0.01 mS/cm from
10.00 to 99.99 mS/cm; 0.1 mS/cm from 100.0 to 400.0 mS/cm; automatic mS/cm: 0.001 mS/cm from 0.000 to 9.999 mS/cm;
0.01 mS/cm from 10.00 to 99.99 mS/cm; 0.1 mS/cm from 100.0 to 400.0 mS/cm±1% of reading or ±1 µS/cm whichever is greater  

automatic single point, with six standard solutions (84 µS/cm, 1413 µS/cm,
5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm) or custom point

TDS Range
Resolution  

Accuracy

Calibration

0.0 to 400.0 ppt (g/L) (the maximum value depends on the TDS factor)
manual: 1 ppm (mg/L); 0.001 ppt (g/L); 0.01 ppt (g/L); 0.1 ppt (g/L); 1 ppt (g/L); automatic: 1 ppm (mg/L) from
0 to 9999 ppm (mg/L); 0.01 ppt (g/L) from 10.00 to 99.99 ppt (g/L); 0.1 ppt (g/L) from 100.0 to 400.0 ppt (g/L);
automatic ppt (g/L): 0.001 ppt (g/L) from 0.000 to 9.999 ppt (g/L); 0.01 ppt (g/L) from 10.00 to 99.99 ppt (g/L);
0.1 ppt (g/L) from 100.0 to 400.0 ppt (g/L)  

±1% of reading or ±1 ppm (mg/L) whichever is greater

based on conductivity or salinity calibration

Resistivity                     Range
Resolution
Accuracy
Calibration
0 to 999999 Ω•cm; 0 to 1000.0 kΩ•cm; 0 to 1.0000 MΩ•cm
dependent on resistivity reading
based on conductivity or salinity calibration
0.00 to 70.00 PSU
Salinity Range
Resolution
Accuracy
Calibration
0.00 to 70.00 PSU
0.01 PSU
±2% of reading or ±0.01 PSU whichever is greater
based on conductivity calibration
Seawater σ Range
Resolution
Accuracy
Calibration
0.0 to 50.0 σt, σ₀, σ₁₅
0.1 σt, σ₀, σ₁₅
±1 σt, σ₀, σ₁₅
based on conductivity or salinity calibration
Temperature Range
Resolution
Accuracy
Calibration
-5.00 to 55.00°C; 23.00 to 131.00°F; 268.15 to 328.15K
0.01°C; 0.01°F; 0.01K
±0.15°C; ±0.27°F; ±0.15K
automatic at one custom point
Additional
Specifications                 
Temperature Compensation
Logging MemoryLogging Interval
PC Connectivity
Environment
Battery Type / Life  

Dimensions / Weight

automatic from -5 to 55°C (23 to 131°F)

45,000 records
(continuous logging or log-on-demand of all parameters)

one second to three hours
via USB (with Hanna PC software)
0 to 50°C (32 to 122°F); RH 100% IP67
1.5V AA batteries (4) / approximately 360 hours of continuous use without backlight (50 hours with backlight)
185 x 93 x 35.2 mm (7.3 x 3.6 x 1.4”) / 400 g (14.2 oz.)

 
Ordering
Information         
HI981954 is supplied with HI7698195/4 multiparameter probe with 4m (13’) cable, HI7698194-1 pH/ORP sensor, HI7698194-3 EC sensor, HI7698295
short protective probe shield, HI9828-20 quick calibration solution, HI76981952 probe maintenance kit, HI7698290 calibration beaker, HI9298194 PC
software, HI920015 micro USB cable, batteries (4), quality certificate, and instruction manual in a rugged carrying case with custom insert.
Accessories HI710034 orange protective rubber boot

 