[:lt]Laboratorinis stalinis pH/mV/ISE bei EC/TDS/druskingumo/varžos matuoklis - HI5522[:en]Laboratory Research Grade Benchtop pH/mV/ISE and EC/TDS/Salinity/Resistivity Meter - HI5522[:] — Laboratory Research Grade Benchtop pH/mV/ISE and EC/TDS/Salinity/Resistivity Meter - HI5522 [:lt]Multiparametrinis pH/mV/ISE/EC/TDS/druskingumo/temp stalinis matuoklis HI2550[:en]Multiparameter pH/mV/ISE and EC/TDS/Salinity Benchtop Meter - HI2550[:] Multiparameter pH/mV/ISE and EC/TDS/Salinity Benchtop Meter - HI2550 —

Autoranging Benchtop EC/TDS/Salinity Meter with ATC – HI2300

505.00  + PVM

The HI2300 is an auto ranging EC/TDS/Salinity and temperature benchtop meter. 

SKU: HI2300-02 Categories: ,

Out of stock

It is no longer produced as of 01.01.2022. Choose HI2003-02 arba HI2030-02 edge meters. 

 

The HI2300 is an auto ranging EC/TDS/Salinity and temperature benchtop meter. The meter utilizes a four-ring potentiometric conductivity probe with platinum sensors to offer versatility over typical amperometric designs. By utilizing the four-ring probe, it is possible to measure very low or very high conductivity levels without having to change probes. The HI2300 also offers three options for temperature compensation, as well as an adjustable temperature coefficient (β) and TDS factor. GLP capabilities and PC connectivity via USB provide users with a simple way to organize and manage their data.

TDS

Autoranging EC and TDS measurements

Cell Constant

Adjustable Cell Constant

ATC

Automatic Temperature Compensation

Details

The HI2300 is a durable benchtop EC/TDS/Salinity and temperature meter that features a four-ring potentiometric probe, one-point calibration, and a USB port for computer connectivity. This meter is auto ranging to choose the appropriate conductivity and total dissolved solids (TDS) range, and can easily be switched to salinity mode to measure from 0.0 to 400.0% NaCl.

Features at-a-glance

Autoranging – In the EC and TDS measurement modes, the HI2300 automatically chooses the optimum measurement scale based on the sample solution to maintain the highest possible accuracy.

Automatic Calibration – EC and TDS are automatically calibrated at one point with a choice of six pre-programmed standards. Salinity is calibrated at one point using the HI7037 100% NaCl standard solution.

Four-ring EC Probe – The HI2300 meter is supplied with the HI76310 platinum, four-ring EC/TDS probe with a built-in temperature sensor that operates over a wide range from 0.00 μS/cm to 500.0 mS/cm*.

Temperature Compensation – Temperature can be compensated for automatically (ATC) or manually (MTC) from -20.0 to 120.0°C, or it can be disabled for actual conductivity or TDS measurements. The temperature correction coefficient, also referred to as β, is adjustable from 0.00 to 6.00 %/°C.

Adjustable TDS Factor – The factor that relates conductivity to total dissolved solids is based on the type of sample being measured. For users to get an accurate determination of TDS based on their unique solution, the TDS factor is adjustable from 0.40 to 0.80.

GLP Data – The calibration data including date, time, standards used, offset and cell constant can be accessed at any time along with the current measurement by selecting the Good Laboratory Practice (GLP) display option.

Data Logging – The log-on-demand feature allows up to 500 data points to be recorded. The data can be recalled on the meter display with a simple press of the “RCL” button.

Data Transfer – Data can be exported to a computer for data review and storage with a USB cable and HI92000 software (both sold separately).

Weight 1.3 kg (2.9 lbs)
EC Range 0.00 to 29.99 μS/cm, 30.0 to 299.9 μS/cm, 300 to 2999 μS/cm, 3.00 to 29.99 mS/cm, 30.0 to 200.0 mS/cm, up to 500.0 mS/cm (absolute EC)**
EC Resolution 0.01 μS/cm, 0.1 μS/cm, 1 μS/cm, 0.01 mS/cm, 0.1 mS/cm
EC Accuracy ±1% of reading (±0.05 μS/cm or 1 digit, whichever is greater)
EC Calibration automatic, one point with six memorized values (84, 1413, 5000, 12880, 80000, 111800 μS/cm)
TDS Range 0.00 to 14.99 ppm (mg/L), 15.0 to 149.9 ppm (mg/L), 150 to 1499 ppm (mg/L), 1.50 to 14.99 g/L, 15.0 to 100.0 g/L, up to 400.0 g/L (absolute TDS)**, with 0.80 conversion factor
TDS Resolution 0.01 ppm, 0.1 ppm, 1 ppm, 0.01 g/L, 0.1 g/L
TDS Accuracy ±1% of reading (±0.03 ppm or 1 digit, whichever is greater)
Temperature Correction Coefficient selectable from 0.00 to 6.00%/°C (EC and TDS only)
EC to TDS Conversion Factor 0.40 to 0.80 (default value is 0.50)
Salinity (% NaCl) Range 0.0 to 400.0% NaCl
Salinity (% NaCl) Resolution 0.1% NaCl
Salinity (% NaCl) Accuracy ±1% of reading
Salinity (% NaCl) Calibration one point, with HI7037 calibration solution (optional)
Temperature Range -20.0 to 120.0 ºC
Temperature Resolution 0.1 °C
Temperature Accuracy ±0.4°C (±0.8°F) (excluding probe error)
EC/TDS Probe HI76310 platinum, four-ring conductivity/TDS probe with internal temperature sensor and 1 m (3.3’) cable (included)
Logging Type log-on-demand
Logging Memory 500 sample
Connectivity opto-isolated USB
GLP yes
Power Supply 12 VDC adapter (included)
Environment 0 to 50°C (32 to 122°F), RH max 95% non-condensing
Dimensions 235 x 222 x 109 mm (9.2 x 8.7 x 4.3”)
Weight 1.3 kg (2.9 lbs.)
Ordering Information HI2300 is supplied with HI76310 conductivity probe, 12 VDC adapter, and instruction manual.
Notes *Uncompensated conductivity (or TDS) is the conductivity (or TDS) value without temperature compensation.