Cyanide Portable Photometer – HI96714

242.00  + PVM

The HI96714 Cyanide Photometer combines accuracy and ease of use in an ergonomic, portable design. 

SKU: HI96714 Category:

Out of stock

This photometer is not produced anymore. HI96714 was changed into new series advanced optical system photometer HI97714C.

 

The HI96714 Cyanide Photometer combines accuracy and ease of use in an ergonomic, portable design. Using the HI93714-0 ready made reagents a user can accurately determine the concentration of cyanide in samples within a 0.000 to 0.200 mg/L (ppm) range. The HI96714 offers many advanced features including the exclusive CAL Check function used for performance verification and calibration of the meter.

CAL Check Validation and Calibration

Built-in Timer

Automatic Shut Off

 

Details

The HI96714 portable photometer is for the measurement of cyanide. Hanna’s portable photometers feature an advanced optical system; the combination of a special tungsten lamp, a narrow band interference filter, and silicon photodetector ensure accurate photometric readings every time. The Hanna exclusive CAL Check feature utilizes ready-made, NIST traceable standards to verify both meter validation and calibration. The exclusive cuvette locking system ensures that the cuvette is inserted into the measurement cell in the same position every time to maintain a consistent path length.

 

Features at-a-glance

CAL Check – Allows for performance verification and calibration of the meter using a CAL Check NIST traceable secondary standard.

GLP – Records the date of the last user calibration performed.

Built in timer – Allows the use of a timer to ensure the appropriate 3 minutes and 30 seconds reaction time is used for the chemical reaction. At the end of the timer countdown the meter will automatically take the reading. This feature ensures consistency across multiple users.

Indexing indent on meter – The meter has an indent that allows for a lock and key fit with the cuvette cap. This ensures that the cuvette is indexed consistently (same position) in order to maintain the same path length for accurate results.

Cooling Lamp Indicator – It is necessary to maintain a consistent temperature of the optical components in order to maintain a narrow wavelength band of light. This photometer has a cooling lamp indicator that is displayed for a short period of time before each measurement in order to allow the components to cool and obtain the highest accuracy possible.

Error messages – Messages on display alerting to problems including no cap, over range and under range readings, and light source error.

Auto-shut off – Automatic shut off after 10 minutes of non-use when the meter is in measurement mode. Prevents wastage of batteries in the event the meter is accidentally not turned off.

Battery status indicator

The instrument has a battery level indicator to display reaming life as follows:

• 3 lines for 100 % capacity

• 2 lines for 66 % capacity

• 1 line for 33 % capacity

• Battery icon blinks when the capacity is less than 10 %.

Units of measure – Appropriate unit of measure is displayed along with reading.

 

The term “cyanide” refers to all of the CN groups in cyanide compounds that can be determined as the cyanide ion, CN. Originating in water primarily from metallurgical and galvanic industrial plants, cyanide is highly toxic to the human nervous system.

The HI96714 uses an adaptation of the Pyridine-Pyrazolone method per the Standard Methods for the Examination of Water and Wastewater to measure cyanide concentrations of less than 0.200 mg/L (ppm). When the pyridine-based reagent is added to samples containing cyanide, the sample will turn a blue hue; the greater the concentration, the deeper the color. The associated color change is then colorimetrically analyzed according to the Beer-Lambert Law. This principle states that light is absorbed by a complementary color, and the emitted radiation is dependent upon concentration. For cyanide determination, a narrow band interference filter at 610 nm (orange) allows only orange light to be detected by the silicon photodetector and omits all other visible light emitted from the tungsten lamp. As the change in color of the reacted sample increases, absorbance of the specific wavelength of light also increases, while transmittance decreases.

 

CAL Check™ standards with certificate

The HI96714-11 CAL Check standards are used for calibration and performance verification of photometers with the CAL Check function.

Supplied with Certificate of Analysis

  • Lot number
  • Expiration date
  • Standard value @ 25 °C
  • Reference meter NIST traceable

Provided storage containers

  • Light tight
  • Protects from accidental breakage

 

 

 
Cyanide Range 0.000 to 0.200 mg/L
Cyanide Resolution 0.001 mg/L
Cyanide Accuracy ±0.005 mg/L ±3% of reading @ 25°C
Cyanide Method Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Pyridine-Pyrazalone method. The reaction between cyanide and reagents causes a blue tint in the sample.
Photometer/Colorimeter Light Source Tungsten Lamp
Photometer/Colorimeter Light Detector Silicon Photocell with narrow band interference filter @ 610 nm
Battery Type/Life 9V battery
Automatic Shut-Off After 10 minutes of non-use in measurement mode; after 1 hour of non-use in calibration mode; with last reading reminder.
Environment 0 to 50°C (32 to 122°F); RH max 95% non-condensing
Weight 360g (12.7oz.)
Dimensions 192 x 104 x 69 mm (7.6 x 4.1 x 2.7”)
Ordering Information HI96714 is supplied with two sample cuvettes with caps, 9V battery, instrument quality certificate, and instruction manual. *CAL Check™ standards and testing reagents sold separately.

 