[:lt]HI1230Y - Geliu užpildytas PEI kūno pH elektrodas su BNC + RCA jungtimi[:en]HI1230Y - Gel Filled PEI Body pH Electrode with BNC + RCA Connector[:] — HI1230Y - Gel Filled PEI Body pH Electrode with BNC + RCA Connector [:lt]HI981304 - pH/EC/TDS Kombinuotas testeris[:en]HI981304 - pH/EC/TDS Combo Tester[:] HI981304 - pH/EC/TDS Combo Tester —

HI1048Y – pH Electrode with Clogging Prevention System (CPS™) and BNC + RCA Connector

273.00  + PVM

The HI1048Y refillable glass-bodied electrode with a sleeve junction and a BNC + RCA connector may become the new food tester’s favorite.

SKU: HI1048Y Category:

Available on backorder - This item may take 2-3 weeks to deliver.

The HI1048Y refillable glass-bodied electrode with a sleeve junction and a BNC + RCA connector may become the new food tester’s favorite. This electrode allows for use of a BNC style probe without a second (separate) NTC thermistor temperature probe. A double junction design paired with Hanna’s Clogging Prevention System (CPS) technology allows for easy testing in samples that can clog traditional electrodes,

  • BNC + RCA connection replaces the need to also have the separate HI7662 temperature probe
  • Clogging Prevention System (CPS™) Technology
  • Perfect for testing in wines, foods, juices and more

Hanna Instruments offers a wide variety of pH electrodes that are designed for many different applications. The type of glass used for sensing pH, bulb shape, body material, type of junction, type of reference and electrolyte used are just some of the design considerations.

The HI1048Y uses low-temperature (LT) glass, spherical bulb, glass body, sleeve junction with CPS™ technology, double junction and is refillable with 3.5M KCl, making it ideal for wine must, juice, and other samples with a high solids content.

 

Low-Temperature Glass Formulation

Low Temperature (LT) glass, as the name implies, is a glass formulation that is used for low-temperature samples and it allows for fast stabilization and more accurate results at lower temperatures. The HI1048Y is suitable to use with samples that measure from -5 to 60°C.

Spherical Glass Tip

The spherical shaped tip design allows for a wide area of contact with the measured sample. This permits a faster electrode response with a higher degree of stability.

CPS Sleeve Junction

Clogging Prevention System (CPS) technology is an innovation in for the improvement of pH measurements in samples that have a high solids content. Conventional pH electrodes use ceramic junctions that can clog quickly when used in samples that have a high solids content such as wine must or juice. When the junction is clogged, the electrode does not function. CPS technology utilizes the porousness of ground glass coupled with a PTFE sleeve to prevent clogging of the junction. The ground glass allows the proper flow of the liquid, while the PTFE sleeve repels solids. As a result, pH electrodes with CPS technology take up to 20 times longer to be fouled as compared to conventional electrodes.

Double Junction Reference

A double junction electrode has an internal compartment surrounding the reference wire. Silver ions are present in the electrolyte of the internal compartment, which houses the Ag/AgCl reference wire; the electrolyte outside this compartment is silver free. The double junction design means that virtually no silver from the electrode enters the sample. This design allows measurement in applications where silver ions in the sample are undesirable or for samples that contain sulfides that can cause the silver to precipitates and clog the junction. The clogging of the junction will result in drifty and erratic readings.

Refillable

The HI1048Y is a refillable probe. Since it is a double junction pH electrode, the fill solution is the HI7082 3.5M KCl. This solution does not contain any silver as with single junction electrode fill solutions.

Glass Electrode Body

The glass body is ideal for laboratory use. The glass is resistant to many harsh chemicals and is easily cleaned. The glass body also allows for a fast transfer of heat to the internal reference electrolyte. The mV generated by the reference cell is temperature dependent. The faster the equilibrium the steadier the reference potential.

BNC + RCA Connector

The HI1048Y has a BNC + RCA connector. This type of connector is unique as it is a BNC electrode that no longer requires the separate HI7662 temperature sensor. The RCA part of the connector incorporates an NTC thermistor temperature sensor. These types of connectors tend to be proprietary. 

Single Junction Versus Double Junction pH Electrodes

Conventional electrodes are normally single junction. As depicted by the figure above, these electrodes have only a single junction between the internal reference wire and the external solution. Under adverse conditions, such as high pressure, high temperature, highly acidic or alkaline solutions, the positive flow of the electrolyte through the junction is often reversed resulting in the ingress of sample solution into the reference compartment. If this is left unchecked, the reference electrode can become contaminated, leading to complete electrode failure. Another potential problem with single junction electrodes is the clogging of the junction due to silver chloride (AgCl) precipitation. Silver can be easily precipitate in samples that contain Tris buffer or heavy metals. When the electrolyte solution makes contact with the sample, some AgCl will precipitate on the external face of the junction. The result is drifty readings obtained from the sensor.

Hanna’s double junction system, as the name implies, has two junctions, only one of which is in contact with the sample as shown in the figure. Under adverse conditions, the same tendency of sample ingress is evident. However, as the reference electrode system is separated physically from the intermediate electrolyte area, the contamination of the electrode is minimized. The likelihood of clogging of the junction is also reduced with a double junction electrode since the outer reference cell uses a fill solution that is “silver-free”. Since there is no silver present, there is no precipitate forming to clog the junction.

Specifications 

Specification Name Detail
SKU HI1048Y
Description pH electrode with CPS (Clogging Prevention System) with BNC + RCA Connector
Reference double, Ag/AgCl
Junction / Flow Rate CPS
Electrolyte KCl 3.5M
Max Pressure 0.1 bar
Range pH: 0 to 12
Recommended Operating Temperature -5 to 60°C (23 to 140°F)
Glass Type LT (low temperature)
Tip / Shape dome (dia: 8 mm)
Temperature Sensor yes
Matching Pin no
Amplifier no
Body Material glass
Cable coaxial
Recommended Use must in wine, food
Connection BNC + RCA
Special Note Compatible with: HI5222, HI5521, HI5522, HI84243, HI84502, HI84529, HI84530, HI84531, HI84532, HI84533, HI84534, HI901W, HI931, HI932

Related products