Accurate and easy to use, the HI9810462 HALO2 Wireless Refillable pH Tester is ideal for laboratory pH measurements in samples that would be a challenge for standard design pH electrodes. This HALO2 can be used as a stand-alone pH tester or can be connected to a smart device with the Hanna Lab App. The Hanna Lab App turns the device into a full-featured pH meter that features live-readings, data sharing, manual temperature compensation, calibration reminder, pH (mV) and temperature alarms. The Hanna Lab App also features Hanna Cloud compatibility.
The HI9810462 features a non-porous, glass body that is easy to clean and withstands harsh chemicals. The double junction reference design presents a silver-free electrolyte solution interacting with the sample, making the electrode less susceptible to clogging and guaranteeing a fast response and stable reading.
The triple ceramic inner junction allows a higher flow rate of electrolyte from the reference cell into the measurement sample. A typical pH electrode will have a single ceramic junction that provides a flow rate of 15 to 20 μL/hour. The HI9810462 has three ceramic junctions in contact with the sample providing for a flow rate of 40 to 50 μL/hour. The increased flow provides a greater continuity between the reference electrode and the indicating electrode, making it suitable for non-homogenous solutions and samples with highly alkaline pH values. Built-in temperature sensor at the tip of the pH electrode allows for rapid determination of the sample temperature and a high-accuracy temperature reading.
- The integrated Bluetooth® module allows the tester to be connected to a compatible smart device with the Hanna Lab App
- Compact, waterproof casing, and automatic pH calibration at up to three points, or five points when used with the Hanna Lab App
- Automatically temperature compensated readings
- Spherical tip
- Provides maximum surface area for accurate measurements.
- Glass body
- Features a non-porous glass body that is easy to clean and withstands harsh chemicals.
- Refillable
- The double junction reference design presents a silver-free electrolyte solution interacting with the sample, making the electrode less susceptible to clogging and guaranteeing a fast response and stable reading.
- Built-in temperature sensor
- Built-in temperature sensor at the tip of the pH electrode allows for rapid determination of the sample temperature and a high-accuracy temperature reading.
Specifications
|
pH
|
Range | 0.00 to 14.00 pH |
| Resolution | 0.01 or 0.1 pH | |
| Accuracy (@25°C/77°F) | ±0.02 pH | |
| Calibration | Up to three points or five points * Automatic buffer recognition with Standard Hanna® buffers (pH 1.68*, 4.01, 7.01, 10.01, 12.45*) or NIST (pH 1.68 *, 4.01, 6.86, 9.18, 12.45*) |
|
|
mV*
|
Range | pH/mV conversion |
| Resolution | 0.1 or 1 mV | |
|
Temperature
|
Range | -5.0 to 80.0°C (23.0 to 176.0°F) |
| Resolution | 0.1 °C; 0.1 °F | |
| Accuracy (@25°C/77°F) | ±0.5 °C; ±0.9 °F | |
| Temperature compensation | Automatic (ATC) or Manual (MTC) * | |
|
Electrode
|
Body material | Glass |
| Glass | High Temperature (HT) | |
| Junction | Triple ceramic | |
| Flow rate | 40 to 50 μL/hour | |
| Reference cell | Double, Ag/AgCl | |
| Electrolyte | 3.5M KCl (refillable) | |
| Tip / Shape | Spheric, Ø 9 mm (Ø 0.35”) | |
| Outer diameter | 12 mm (0.47”) | |
| Length | 120 mm (4.7”) | |
| Battery type | 3V Lithium – CR2032 | |
| Battery life | Approximately 1000 hours (500 hours with Bluetooth enabled) | |
| Environment | 0 to 50 °C (32 to 122 °F) | |
| IP rating | IP65 | |
| Dimensions / Weight | 51 x 206 x 21 mm (2.0 x 8.1 x 0.8“) / 60 g (2.1 oz.) | |