GroLine Monitor – HI981421

511.50  + PVM

GroLine Monitor is designed to offer you the combination of pH, conductivity (or total dissolved solids) with temperature measurements.

SKU: HI981421 Category: Tag:

Available on backorder - This item may take 2-3 weeks to deliver.

GroLine Monitor is designed to offer you the combination of pH, conductivity (or total dissolved solids) with temperature measurements. All operations and settings, including calibration buffers and temperature scale selection, are included in a easy to use Monitor setup.
 
The supplied HI1285‑8 (with HI981420) or HI1285‑9 (with HI981421) multiparameter probe measures pH, EC/TDS and temperature in one convenient, rugged probe. A solid‑state preamplifier is integrated into the probe to protect the pH measurement from transient electrical noise. Sources of electrical noise include ballasts used in lighting and pumps to circulate water and nutrient solutions. If pH reading in pH 7.01 buffer is around 4.01, the probe is broken.
 
Measurements are automatically logged in the GroLine Monitor internal memory every 15 minutes and can be viewed using the plot/record log feature or exported to a Flashdrive, or by connecting to a PC. The three main operating modes of this Monitor are measurement, setup and logging. 
 
Other features include:
• can be integrated into a fertilizer system
• pH/EC/TDS probe with built‑in temperature sensor
• as afailsafe feature, if the pH bulb breaks, the pH will be around 4.01
• multiple viewing options for on‑screen data
• quick calibration using Quick cal solution for pH and EC simultaneously
• pH and EC calibration timeout reminder (1‑30 days)
• visual plot screens for real‑time data and log history
• 24 hours summary screens (plot and details)
• alarms feature for all parameters
• automatic continuous logs, 15 minutes intervals, 30 days storage
• logged data can be stored and later exported to a USB C drive or transferred to a PC using USB C connection
• GLP feature calibration data for pH and EC (up to 5 calibrations)
• dedicated help with contextual message
• setup, calibration and operating parameters are stored in nonvolatile memory (important to note that if power is lost all settings are retained)
• probe inline mounting for fast measurement feedback (HI1285‑9 only)
• ambient light sensor for ON/OFF backlight

 

pH specifications 

Range 0.00 to 14.00 pH, 0.0 to 14.0 pH*
Resolution 0.01 pH, 0.1 pH 
Accuracy (25ºC/77 ºF) ±0.05 pH, ±0.1 pH
Calibration

one or two‑point calibration (using pH 4.01, 7.01, 10.01 buffers) using auto buffer recognitionone‑point calibration using quick calibration solution

Temperature compensation  automatic: 0.0 to 60.0 °C; 32.0 to 140.0 °F

EC specifications 

Range 0.00 to 10.00 mS/cm
Resolution 0.01 mS/cm
Accuracy (25ºC/77 ºF) ±0.10 mS/cm (0.00 to 5.00 mS/cm)    ±0.20 mS/cm (5.00 to 10.00 mS/cm)
Calibration

one‑point at 1.41 mS/cm or 5.00 mS/cm using auto standard recognition.one‑point calibration using quick calibration solution

Temperature compensation (not user selectable) automatic, with b = 1.9%/°C

TDS specifications

 
Range

0 to 5000 ppm (0.5 TDS Factor)**
0 to 7000 ppm (0.7 TDS Factor)**

Resolution 10 ppm
Accuracy (25ºC/77 ºF) ±2% FS
Calibration through EC calibration
TDS factor 0.5 (500 ppm) or 0.7 (700 ppm)
Note: pH range will be reduced to actual probe specification**
Note: 1000 μS/cm = 500 ppm with 0.5 TDS Factor = 700 ppm with 0.7 TDS Factor

 

Temperature specifications

Range

0.0 to 60.0 °C / 32.0 to 140.0 °F

Resolution 0.1 °C / 0.1 °F
Accuracy (25ºC/77 ºF) ±0.5 °C /±1.0 °F

Additional specifications

Automatic logging

Measurements (pH, EC, TDS, temperature) Min/Max/average/status logged continuously at 15 minutes interva

Recall graphic modes
Export on USB‑C flash drive or PC
Log files in CSV format
30 days stored data at 15 minutes interval
Display 128 x 64 pixel B/W LCD with green backlight
Automatic backlight dimming using ambient light sensor
GLP Good Laboratory Practice with last 5 pH and EC calibration history
Monitor protection rating IP65 (dust and low pressure water jets
Alarms High and Low with enable/disable option for all parameters
USB‑C (Host/Device) Export logged data on USB flash drive / PC
Power supply 12VDC adapter (included)
Environment

0 to 50 °C (32 to 122 °F), RH max 95% non‑condensing

Monitor dimensions 125 x 185 x 38 mm (4.92 x 7.28 x 1.49”)
Monitor weight 333 g (11.7 oz.)
Ordering information

HI981420   is   supplied   with:   HI1285‑8   multiparameter   probe,   HI50036  Quick  cal  buffer  solution  (2  sachets),  HI700661  pH  electrode  cleaning  solution  for  agriculture  (2  sachets),  DC  power  adapter  (1)  and  instruction  manual.  HI981420‑01 withUS  plug,  HI981420‑02 with EU plugHI981421   is   supplied   with:   HI1285‑9   inline   multiparameter   probe, HI50036  Quick  cal  buffer  solution  (2  sachets),  HI700661pH electrode cleaning solution for agriculture (2 sachets), DC power adapter  (1)  and  instruction  manual.  HI981421‑01 withUS  plug,  HI981421‑02 with EU plug

Warranty

GroLine Monitor is warranted for a period of 2 years

 