[:lt]pH/ORP/temperatūros stalinis matuoklis su kalibravimo patikra - HI3220[:en]pH/ORP/Temperature Bench Meter with Calibration Check - HI3220[:] — pH/ORP/Temperature Bench Meter with Calibration Check - HI3220 [:lt]Dviejų išvesčių pH/ORP/ISE/temperatūros stalinis matuoklis su kalibravimo patikra - HI3222[:en]Dual-Input pH/ORP/ISE/Temperature Bench Meter with Calibration Check - HI3222[:] Dual-Input pH/ORP/ISE/Temperature Bench Meter with Calibration Check - HI3222 —

Dual-Input pH/ORP/ISE/EC/Resistivity/TDS/NaCl/Temperature Bench Meter with Calibration Check – HI3512

566.00  + PVM

HI3512 is a dual-input professional pH/ORP/ISE/EC/Resistivity/TDS/NaCl/Temperature bench meter with a graphic LCD designed to provide high accuracy and ease of use both in the laboratory as well as in harsh industrial conditions. This instruments is a dual-input model measuring pH, ORP, or ISE and temperature on channel 1 and EC, Resistivity, TDS, and Salinity (utilizing an EC probe) on channel 2. A Relative mV feature is also available. 

Available on backorder - This item may take 2-3 weeks to deliver.

HI3512 is a dual-input professional pH/ORP/ISE/EC/Resistivity/TDS/NaCl/Temperature bench meter with a graphic LCD designed to provide high accuracy and ease of use both in the laboratory as well as in harsh industrial conditions. This instruments is a dual-input model measuring pH, ORP, or ISE and temperature on channel 1 and EC, Resistivity, TDS, and Salinity (utilizing an EC probe) on channel 2.

This meter features Hanna’s exclusive Calibration Check™, a diagnostics system that ensures accurate readings every time. Calibration Check system eliminates erroneous readings due to dirty or faulty pH electrodes or contaminated buffer solution by alerting users of potential problems during the calibration process.

Throughout the calibration process, users are guided step-by-step by the on-screen tutorial. After calibration, a probe condition indicator is displayed on the LCD informing the user of the overall electrode status.

A variety of interactive user support is available before, during and after measurement. Onscreen tutorials guide users through set-up, calibration and measurement while context sensitive help of any screen is available at a push of a button. The HELP screen accessed by a dedicated HELP button, includes language specific assistance for menu parameters, calibration, log, contact information and accessories for your instrument.

 

Main Features:

  • Dual-input channel
    • pH/mV/ISE and temperature measurements (Channel 1)
    • EC/TDS, NaCl/Resistivity and temperature measurements (Channel 2)
  • Up to 5 point pH calibration with 7 standard buffers and 2 custom buffers to choose from
  • Up to two points EC calibration with 7 memorized standards for calibration to choose from
  • Messages on the graphic LCD for an easy and accurate calibration
  • Diagnostic features to alert the user when the electrode needs cleaning
  • Relative mV measurements
  • Log on demand up to 400 samples
  • Log interval with log on stability feature up to 600 records
  • Auto Hold feature, to freeze first stable reading on the LCD
  • GLP, last calibration data for pH, Rel mV, ISE, EC, or NaCl
  • PC interface

Primary Applications:

  • Laboratory, Industrial

 

Specifications:

pH
Range –2.0 to 20.0 pH
–2.00 to 20.00 pH
–2.000 to 20.000 pH
Resolution 0.1 pH
0.01 pH
0.001 pH
Accuracy ±0.01 pH
±0.002 pH
Calibration Up to 5 points
7 standard buffers (1.68, 4.01, 6.86, 7.01, 9.18,
10.01, 12.45)
2 custom buffers
ORP
Range ±2000.0 mV
Resolution 0.1 mV
Accuracy ±0.2 mV
ISE
Range 1.00 E-7 to 9.99 E10 conc.
Resolution 3 digits 0.01, 0.1, 1, 10 conc.
Accuracy ±0.5% of reading (monovalent ions)
±1% of reading (divalent ions)
Calibration Up to 5 points
6 standards (0.1, 1, 10, 100, 1000, 10000 ppm)
Temperature Channel 1
Range –20.0 to 120.0 °C (–4.0 to 248.0 °F)
Resolution 0.1 °C (0.1 °F)
Accuracy ±0.2 °C (±0.4 °F) (excluding probe error)
Rel mV offset range ±2000 mV
Slope calibration From 80 to 110%
pH Temperature compensation Manual
Automatic
Electrode Hanna BNC connection pH / ORP / ISE electrodes
Temperature probe RCA connection
Recommended option: HI7662-TW
EC
Range 0 to 400 mS/cm
(shows values up to 1000 mS/cm)
Actual conductivity 1000 mS/cm
0.001 to 9.999 μS/cm
10.00 to 99.99 μS/cm
100.0 to 999.9 μS/cm
1.000 to 9.999 mS/cm
10.00 to 99.99 mS/cm
100.0 to 999.9 mS/cm
1000 mS/cm
(autoranging)
Resolution 0.001 μS/cm
0.01 μS/cm
0.1 μS/cm
0.001 mS/cm
0.01 mS/cm
0.1 mS/cm
1 mS/cm
Accuracy ±1% of reading (±0.01 μS/cm or 1 digit whichever
greater) excluding probe error
Calibration Up to 2 points
7 standards (0.00 μS/cm, 84.0 μS/cm, 1.413 mS/
cm, 5.00 mS/cm,12.88 mS/cm, 80.0 mS/cm, 111.8
mS/cm)
Resistivity
Resolution 1.0 to 99.9 ohms
100 to 999 ohms
1.00 to 9.99 Kohms
10.0 to 99.9 Kohms
100 to 999 Kohms
1.00 to 9.99 Mohms
10.0 to 100.0 Mohms
(autoranging)
Accuracy 0.1 ohm
1 ohm
0.01 Kohms
0.1 Kohms
1 Kohms
0.01 Mohms
0.1 Mohms
Calibration ±1% of reading (±10 ohms or 1 digit whichever
greater) excluding probe error
TDS
Range 0.000 to 9.999 ppm
10.00 to 99.99 ppm
100.0 to 999.9 ppm
1.000 to 9.999 g/L
10.00 to 99.99 g/L
100.0 to 400.0 g/L
(autoranging)
Resolution 0.001 ppm, 0.01 ppm, 0.1 ppm
0.001 g/L, 0.01 g/L, 0.1 g/L
Accuracy ±1% of reading (±0.05 ppm or 1 digit whichever
greater) excluding probe error
TDS Factor 0.40 to 1.00
Salinity
Range % NaCl: 0.0 to 400.0 %
Resolution 0.1%
Accuracy ±1% of reading excluding probe error
Temperature Channel 2
Range −20.0 to 120.0 °C (4.0 to 248.0 °F)
Resolution 0.1 °C / 0.1 °F
Accuracy ±0.2 °C (excluding probe error)
Cell constant setup 0.010 to 10.000
NaCl Calibration Max. 1 point only (with HI7073 standard)
EC probe DIN connection
Recommended option: HI76310
Temperature source Automatic
Manual
EC temperature
compensation
NoTC, MTC, ATC
Reference temperature 15, 20, 25 °C
Temperature coeficient 0.00 to 10.00 %/°C
Log on demand 400 samples
Lot logging 5, 10, 30 seconds
1, 2, 5, 10, 15, 30, 60, 120, 180 minutes, AutoEnd
(maximum 600 samples)
Environment 0 to 50 °C (32 to 122 °F)
max. RH 55% non-condensing
Power supply 12 Vdc power adapter
PC interface opto-isolated USB
Dimensions 235 x 207 x 110 mm (9.2 x 8.14 x 4.33”)
Weight (meter only) 1.8 Kg (4.1 lb)

 

Ordering Information:

HI3512-01 (115V) and HI3512-02 (230V) is supplied with:

  •  Benchtop meter
    • HI1131B pH electrode
    • HI76310 Conductivity probe
    • HI7662-TW Temperature probe
    • HI76404W Electrode holder
    • pH 4.01 buffer solution, 20 mL sachet
    • pH 7.01 buffer solution, 20 mL sachet
    • Cleaning solution, 20 mL, 2 sachets
    • Electrolyte solution
    • Capillary pipette
    • 12 Vdc power adapter
    • Instrument quality certificate with quick reference guide
    • Electrode | probe quality certificates with quick reference guides

Accessories:

(*denotes preferred accessories for web when space is a concern)
Please refer to instruction manual for full accessory list.

pH Buffer Solution

HI5016 pH 1.68 Buffer Solution, 500 mL bottle
*HI5004 pH 4.01 Buffer Solution, 500 mL bottle
HI5068 pH 6.86 Buffer Solution, 500 mL bottle
*HI5007 pH 7.01 Buffer Solution, 500 mL bottle
HI5091 pH 9.18 Buffer Solution, 500 mL bottle
*HI5010 pH 10.01 Buffer Solution, 500 mL bottle
HI5124 pH 12.45 Buffer Solution, 500 mL bottle

pH Electrodes

HI1043B Glass-body, double junction, refillable, combination pH electrode.
Use: strong acid/alkali.
HI1053B Glass-body, triple ceramic, conic shape, refillable, combination pH electrode.
Use: emulsions.
*HI1083B Glass-body, micro, Viscolene, non-refillable, combination pH electrode.
Use: biotechnology, micro titration.
*HI1131B Glass-body, double junction, refillable, combination pH electrode.
Use: general purpose.
HI1330B Glass-body, semimicro, single junction, refillable, combination pH electrode.
Use: laboratory, vials.
HI1331B Glass-body, semimicro, single junction, refillable, combination pH electrode.
Use: flasks.
*HI1230B Plastic-body (PEI), double junction, gel-filled, combination pH electrode.
Use: general, field.
FC911B Plastic-body (PVDF), double junction, refillable with built-in amplifier, combination pH
electrode.
Use: very high humidity.
*HI1413B Glass-body, single junction, flat tip, Viscolene, non-refillable, combination pH
electrode.
Use: surface measurement.

ORP Electrodes

*HI3230B Plastic-body, gel-filled, combination platinum ORP electrode. Use: general purpose.

ORP Pretreatment Solutions

HI7091L Reducing Pretreatment Solution, 500 mL bottle +14 g (set)
HI7092L Oxidizing Pretreatment Solution, 500 mL bottle

Electrode Cleaning Solution

*HI7061L General Purpose Solution, 500 mL bottle
Electrode Storage Solution
*HI70300L Storage Solution, 500 mL bottle

EC Probe

*HI76310 Platinum, four-ring EC/TDS probe with a built-in temperature sensor and DIN
connection

EC Calibration Solutions

HI6033 84 μS/cm, 500 ml, bottle
HI6031 1413 μS/cm, 500 ml, bottle
HI7039L 5000 μS/cm, 500 ml, bottle
HI7030L 12880 μS/cm, 500 ml, bottle
HI7034L 80000 μS/cm, 500 ml, bottle
HI7035L 111800 μS/cm, 500 ml, bottle
HI7037L 100% NaCl, 500 ml, bottle

 


Related products