The HI96705-11 Silica CAL Check™ standards provide a simple solution to calibrating and validating the HI96705 Silica Low Range Portable Photometer. This high quality set of standards is manufactured in our state-of-the-art facility and comes supplied with a certificate of analysis. The certificate of analysis provides the lot number, reference values, and expiration date for traceability when certifying the HI96705 photometer.
Supplied with Certificate of Analysis
Standard value of 1.00 ±0.01 mg/L (ppm) SiO2
Reusable with long shelf life stability
Details
HI96705-11 is a high quality set of calibration standards that allows users to validate and calibrate the HI96705 Silica Portable Photometer. The HI96705-11 includes one sealed “A” cuvette for zeroing the meter and one “B” cuvette that is used to validate or calibrate. The standard value of the “B” cuvette is 1.00 ± 0.01 mg/L (ppm) SiO2. An acceptable reading when validating the HI96705 with the standards is 1.00 ±0.05 mg/L @ 25°C. The calibration set is produced in our state-of-the-art manufacturing facility using reagent grade chemicals in a temperature controlled environment.
Supplied with Certificate of Analysis
- Product code
- Lot number
- Expiration date
- Standard value @ 25°C
- Reference meter NIST traceable
Provided storage containers
- Light tight
- Protects from accidental breakage
Validation and calibration of the HI96705 Silica Portable Photometer is quick and easy with the HI96705-11 CAL Check™ Standards.
Package | Sealed standards: 0.00 and 1.00 mg/L |
---|---|
Quantity | 1 set |