[:lt]Skaitmeninis pH elektrodas - HI11311[:en]Digital Glass Body pH Electrode with Matching Pin for General Purpose[:] — Digital Glass Body pH Electrode with Matching Pin for General Purpose [:lt]HI12301 Skaitmeninis pH elektrodas su "Matchin Pin" (edge)[:en]HI12301 Digitale pH-Elektrode mit Matchin Pin, plastic body, Gel (edge)[:] HI12301 Digitale pH-Elektrode mit Matchin Pin, plastic body, Gel (edge) —

Digital PEI Body pH Electrode for Field Applications

136.00  + PVM

HI12300 SMART pH elektrodas,gelio užpildu, PEI korpuse, su integruotu temperatūros jutikliu (edge)

SKU: HI12300 Category:

Available on backorder - This item may take 2-3 weeks to deliver.

The HI12300 is a plastic body, double junction, gel filled pH electrode with a built-in temperature sensor for temperature compensated measurements in a single probe design. This probe connects to the meter with a 3.5 mm connector. The 3.5 mm connector is proprietary to meters that use digital electrodes.

The HI12300 electrode features a built-in microchip that stores sensor type, serial number, and calibration information. It also features a single ceramic junction and the spherical pH sensing portion is made with low temperature glass. The electrode has a rugged PEI (polyetherimide) body that has the chemical resistance of a glass body in an unbreakable plastic making it ideal for field and industrial applications.

digital electrode

Built-In Microchip

 

Double Junction Reference

Double Junction Reference

 

3.5mm Connector

3.5 mm Connector

Details

Built-In Microchip

The built-in microchip stores sensor type, serial number and calibration information including date, time, offset, slope, probe condition and buffers used. This information is automatically retrieved by edge®once the electrode is plugged in. The ability to transfer information allows for hot swapping of probes without having to recalibrate. 

All pH measurements are performed within the electrode and transferred digitally to the meter. This overcomes any noise issues associated with the traditional high impedance analog measuring system. Electrical noise can be generated from a built in temperature sensor and while working in a humid environment.

 

Spherical Glass Tip

The spherical shaped tip design allows for a wide area of contact with the sample. This permits a faster electrode response with a higher degree of stability. The tip uses a glass formulation that has a resistance of approximately 100 megaohms. This means that measuring conditions are optimal around ambient temperatures. Use of high or low temperature glass around ambient temperatures would not be optimal due to a long response time or a shortened lifespan of the electrode.

 

Double Junction Reference

A double junction electrode has an internal compartment surrounding the reference wire. Silver ions are present in the electrolyte of the internal compartment, which houses the Ag/AgCl reference wire; the electrolyte outside this compartment is silver free. The double junction design means that virtually no silver from the electrode enters the sample. This design allows measurement in applications where silver ions in the sample are undesirable or silver precipitates on the junction are likely to form.

 

PEI Electrode Body

The PEI (polyetherimide) electrode body is suitable for a wide range of applications and excels in field measurements due to its durability. The shielding around the spherical glass tip minimizes breakage due to accidental bumping or dropping of the electrode. The PEI plastic is a high quality plastic that is chemically resistant to many harsh chemicals.

 

Built-In Temperature Sensor

HI12300 features a built-in thermistor temperature sensor that is in the tip of the indicating pH electrode. A thermistor temperature sensor provides for a high accuracy temperature reading and should be as close as possible to the indicating pH electrode in order to compensate for the effect that temperature has on the membrane potential. By having an accurate reading it is possible to provide an accurate temperature compensated reading.

 

 

Measurement Range pH: 0 to 13
Reference Cell Type double
Junction Type/Flow Rate ceramic, single / 15-20 μL/h
Type of Electrolyte gel
Body Material PEI
Tip / Shape spheric (dia: 7.5 mm)
Matching Pin no
Max. Pressure 2 bar
Temperature Operating Range -5 to 70°C (23 to 158°F) – LT
Body Length/Overall Length 120 mm / 163.5 mm
Temperature Sensor yes
Outer Diameter 12 mm
Cable 1 m (3.3′)
Connector Type 3.5 mm connector
Amplifier no
Recommended Use field applications

Additional information

Weight 0.319 kg