[:lt]HI12303 - Geliu užpildytas PEI korpuso pH elektrodas su greito prijungimo DIN jungtimi[:en]HI12303 - Gel Filled PEI Body pH Electrode with Quick Connect DIN Connector[:] — HI12303 - Gel Filled PEI Body pH Electrode with Quick Connect DIN Connector HI1330P - pH Electrode for Vials and Test Tubes with BNC + PIN Connector —

HI1144B – Refillable pH Electrode with Calomel References and BNC Connector

200.00  + PVM

The HI1144B is a glass body, refillable, single junction pH electrode with a BNC connector.

SKU: HI-1144B Categories: ,

Available on backorder - This item may take 2-3 weeks to deliver.

The HI1144B is a glass body, refillable, single junction pH electrode with a BNC connector. This electrode has a single ceramic junction in the outer reference cell and the spherical pH sensing portion is made with high temperature glass.  The mercury/mercury (I) chloride (Hg/Hg2Cl2) reference system of the HI1144B provides a stable reference potential without the presence of silver. This design consideration is ideal for applications involving Tris buffer.

  • Calomel References
  • Refillable Electrode
  • High Temperature (HT) Glass

Specifications

Specification Name Detail
SKU HI1144B
Description refillable, combination pH electrode with calomel references
Reference single, Hg/Hg₂Cl₂
Junction / Flow Rate ceramic / 15-20 µL/h
Electrolyte KCl 3.5M
Max Pressure 0.1 bar
Range pH: 0 to 14
Recommended Operating Temperature 0 to 60°C (32 to 140°F) — HT
Tip / Shape spheric (dia: 9.5 mm)
Temperature Sensor no
Amplifier no
Body Material glass
Cable coaxial; 1 m (3.3′)
Recommended Use tris buffer
Connection BNC

Related products