[:lt]pH elektrodas su mikro lempa ir BNC jungtimi - HI1083B[:en]pH Electrode with Micro Bulb and BNC Connector - HI1083B[:] — pH Electrode with Micro Bulb and BNC Connector - HI1083B [:lt]Prailginto ilgio pH elektrodas su Micro lempute ir BNC jungtimi - HI1093B[:en]Extended Length pH Electrode with Micro Bulb and BNC Connector - HI1093B[:] Extended Length pH Electrode with Micro Bulb and BNC Connector - HI1093B —

pH Electrode with Micro Bulb and BNC + Pin Connector – HI1083P

423.00  + PVM

The HI1083P is a glass body, single reference junction micro pH electrode with a BNC and pin connector. 

SKU: HI1083P Category:

Available on backorder - This item may take 2-3 weeks to deliver.

The HI1083P is a glass body, single reference junction micro pH electrode with a BNC and pin connector. The pin connector is used to enable the CAL Check function of the HI2221, HI2222, and HI2223 benchtop meters. This electrode features a bulb tip that is only 3 mm in diameter and can be used to measure pH in samples as little as 100 µL.

This probe is ideal for 96 well plates and for expensive samples that offer little volume to work with. The HI1083P has a unique open junction design in which there is a viscolene (hard gel) electrolyte layer that is free of silver chloride (AgCl) between the sample to be measured and the internal reference cell.

Open Junction 

Micro Bulb Tip

Viscolene Gel Electrolyte 

Details

Hanna Instruments offers a wide variety of pH electrodes that are designed for many different applications. The type of glass used for sensing pH, bulb shape, body material, type of junction, type of reference and electrolyte used are just some of the design considerations.

The HI1083P uses general purpose (GP) glass, micro bulb, glass body, and open junction with viscolene gel electrolyte.

Glass

General Purpose Glass Formulation

General Purpose (GP) glass, as the name implies, is a standard glass formulation that is used for general use. A pH electrode with GP glass will have a resistance of 100 megaohms at 25°C and is suited for measuring pH of samples that are at ambient temperatures. The HI1083P is suitable to use with samples that measure from 0 to 50°C.

Micro Bulb

Micro Bulb Tip

The 3 mm micro bulb tip design allows for measurement of pH in samples with very little volume or space. The HI1083P can be used in samples with volumes as little as 100 µL. It is an ideal probe for measuring the pH in 96 well plates.

Glass Body

Glass Body

The glass body is ideal for laboratory use. The glass is resistant to many harsh chemicals and is easily cleaned. The glass body also allows for a fast transfer of heat to the internal reference electrolyte. The mV generated by the reference cell is temperature dependent. The faster the electrode reaches equilibrium, the steadier the reference potential.

Open Junction

Open Junction Reference

Suspended solids and proteins found in food products will clog a conventional ceramic reference junction. This clogging will impede the measurement circuit between the indicating electrode and the internal reference resulting in slower response time, erratic readings, and frequent electrode replacement. The open junction design consists of a solid gel interface (viscolene) between the sample and internal Ag/AgCl reference. This interface not only prevents silver from entering the sample, but also makes it impermeable to clogging, resulting in a fast response and stable readings.

BNC pin connector

 BNC + Pin Connector

The HI1053P has a BNC and pin connector. The BNC connector is universal in that it can be used on any pH meter that has the female BNC probe input. The pin connector is used to enable the CAL Check function on the following benchtop pH meters; HI122, HI123, HI221, HI222, HI223, HI2221, HI2222, and HI2223.

 

Measurement Range pH: 0 to 13
Reference Cell Type double, Ag/AgCl
Junction Type/Flow Rate open
Type of Electrolyte viscolene
Body Material glass
Tip / Shape spheric (dia: 3 mm)
Matching Pin no
Max. Pressure 0.1 bar
Temperature Operating Range 0 to 50°C (32 to 122°F) – GP
Body Length/Overall Length 120 mm / 163.5 mm
Temperature Sensor no
Outer Diameter 3 mm
Cable coaxial; 1 m (3.3’)
Connector Type BNC + pin
Amplifier no
Recommended Use NMR tubes

Additional information

Weight 0.116 kg